Compact, economical, and easy-to-use, the SAVe II+ is built for dependable use during a crisis. Medical personnel have relied on AutoMedx products devices for over a decade. The new SAVe II+ delivers up to 50% more flow than the SAVe II and can generate up to 20 cmH2O of PEEP. The SAVe II+ replaces the SAVe II.
This kit includes the following:
SAVe II+ Product & Performance Specifications
General:
Dimensions Ventilator 6.5" x 6.25" x 2.0" (81 cubic inches)Weight Ventilator 2.8 lbs. (1.3 kg)Weight of Full Kit 9.9 lbs. (4.5 kg)Operating Mode A/C-CMVMode Description Time cycled, pressure limited, volume targeted.PRESET Heights 4' 3" - 6' 3" in 3-inch incrementsPRESET Target 6 ml/kg of IBW or 30 cmH2O of Working PressureFlow Rate 0 - 40 LPMI:E Ratio 1:2 FixedRespiratory Rate 8 - 30 BPMTidal Volume 200 - 800 mlPeak Inspiratory Pressure 10 - 60 cmH2OPEEP 0 - 20 cmH2OInadvertent PEEP < 2.0 cmH2OFiO2 Range: 21 - 100%Oxygen Input: 0 - 12 LPM from Flow Regulated SourceInspiratory Trigger 2.0 cmH2OInspiratory Resistance < 6 cmH2O @ 30 LPMExpiratory Resistance < 6 cmH2O @ 30 LPMDead Space < 100 mlBattery Duration 8.25-9.25 Hours (see below note)Power Input 100 – 240 VAC / 50-60 HzAudible Alarms Meets 60601-1-8Shock & Vibration IEC 60068-2-6 and 60068-2-36Ingress Protection Vent IP24Protective Case IP67SETTINGS:
Breath Rate (BPM) 8 – 30Tidal Volume (ml) 200 - 800Pressure Limit (cmH2O) 10 - 60PEEP (cmH2O) 0 - 20MEASUREMENTS:
PIP (cmH2O) 0 - 60PEEP (cmH2O) 0 - 25FIXED PARAMETERS:
I:E Ratio 1:2Inspiratory Time (Ti) 0.67 - 2.50 secondsExpiratory Time (Te) 1.3 - 5.0 secondsFlow Rate (Avg LPM) 4 - 40Product Life:
Control Unit 8 yearsBreathing Circuit 3 year shelf life (single use)Notes:
1. All specifications include a tolerance of ±10% of nominal value unless stated otherwise in the Operator’sManual. Test conditions available upon request.2. Delivered Tidal Volume may be materially affected by very low lung compliance (< 0.01 L/cmH2O)3. With a fully charged new battery the 5’9” preset will run approximately 8.25 hrs with (2600 mAh battery) or 9.25 hrs with 2800 mAh battery)